ThermoFisher Fiberlite™ F14-6 x 250y

Manufacturer
ThermoFisher Scientific
Platform
Thermo Scientific™ Sorvall™ RC 6 Plus and Evolution RC superspeed Centrifuges
Type
  • Fixed Angle
Angle
23°
Tube Number
6
Tube Volume
250.0 ml
Tube Diameter
62.0 mm
Tube Length
135.0 mm
Max. Capacity
250 ml
rMAX
137.0 mm
rMIN
37.0 mm
Max. Speed
14,000 rpm
Max. RCF
30,100 x g
k Factor
1,690
Outside Source
Vendor Specifications Catalog No.
ThermoFisher Scientific For use with Thermo Scientific™ Jouan KR22i Centrifuges 096-062128
ThermoFisher Scientific For use with Beckman™ J2 Series Centrifuges 096-062038
ThermoFisher Scientific For use with Hitachi™ Himac CR21/CR22 Series Centrifuges 096-062132
ThermoFisher Scientific For use with Beckman™ Avanti J Series (except the Avanti J-HC) Centrifuges 096-062107
ThermoFisher Scientific For use with Thermo Scientific™ Sorvall™ RC 6 Plus and Evolution RC superspeed Centrifuges 78500

RCF and k-Factor Calculator 

12000 rpm
rpm
At 12000 rpm, the RCF equals 22,056 x g at rMAX and 0 x g at rAV.
The k Factor (adjusted) is 2,300.

RPM Calculator (from RCF) 

10000 x g
x g
To achieve an RCF of 10000 x g at rMAX, set the speed to 8,080 rpm.
To achieve an RCF of 10000 x g at rAV, set the speed to rpm.

Centrifugation Time at Max. Speed (to match reference rotor based on k factors) 

rpm
min.
To use the k Factor to calculate the "Time at Max Speed" needed to match the centrifugation with another rotor, enter the "Reference Rotor," "Speed" and "Time."

Centrifugation Time at Specified Speed (to match reference rotor based on k factors) 

rpm
min.
ThermoFisher Fiberlite™ F14-6 x 250y
12000 rpm
rpm
To use k factors to calculate the "Time" at a "Specified Speed" needed to match the separation with another rotor, enter the "Reference Rotor," "Speed" and "Time." and the "Specified Speed."